Measuring a Reaction Rate- Experiment C
Start up procedure
· Go to “Start”, “All programs”. Chemistry Applications”, and then click “Logger Pro”
Choose the Settings
· Click on “Experiment” and then click “Data Collection…”
· Choose the following options in the dialog box that appears, and then click “Done”
· To choose the correct wavelength settings click on the “Configure Spectrometer” button
· Choose the following options in the dialog box that appears and then click “ok”
Calibrating the Spectrophotometer
· Click on “Experiment” and then click “calibrate; spectrometer:1”
· To make the blank add 0.5mL DMSO and 2.5mL Papain to a clean and dry cuvette. Place parafilm on top and invert four times, then place the blank in the sample holder
· Click “Finish calibration” in the calibration dialog box
· Once calibrated click “ok” and remove the cuvette from the sample holder
Collect Absorbance Values and Generate a Time vs. Absorbance Graph
· In a clean and dry test tube add 2.5mL Papain
· In that same test tube add 0.5mL of either 0.0025M BAPNA or 0.0050M BAPNA
· Place parafilm on top and invert the test tube four times
· Transfer the contents of the test tube into a clean and dry cuvette so that it is 3/4th full
· Wipe the outside of the cuvette with a kimwipe and place in the sample holder
· Click “collect” . It should take no longer than 10 seconds to start collecting data from the time the BAPNA is added to the test tube containing the Papain
· Wait for the program to automatically stop at 50 seconds
· Click “linear fit” , title, save and print the graph
· The graph should look similar to the one shown below
Collect Absorbance Spectra of the Final Product